L Water pressure: 100~392kPa, and the limit peak value is not more than 800kPa (Note: the outlet pressure of most pure water machines is normal pressure, so it is necessary to provide pressurization equipment from the water dispenser supplier or use the water inlet module provided by Mindray Company for pressurization);
L Flow: the continuous flow is not less than 30 liters/hour, and the peak flow is not more than 1 liter/minute. For a water machine with a nominal flow rate of 20 liters/hour, a water storage container of 40 liters or more must be built in;
L water quality: the water supply quality meets the CAP grade water requirements of the American Society of Clinical Pathology, and the specific resistance is not less than 0.5 (mω nan @ 25℃);
Below BS-380, there is no requirement for water pressure.